Definition of scientific term
Chinese name: electric blast drying oven
English name: drying oven on forced convection
Definition: An electric drying oven that uses a ventilator to force air convection inside the tank.
Applied disciplines: mechanical engineering (a subject); laboratory instruments and devices (two subjects); climate environmental testing equipment - climate environmental testing equipment equipment name (three subjects)
Electric blast drying oven box structure:The design is perfect, the box body is machined by CNC machine tool, and the box door adopts the upper open type, which is easy to operate.
The inner liner is made of SUS304 stainless steel plate, and the outer casing is sprayed with A3 plate, which is more smooth and beautiful.
The door system is opened on the side of the circuit system for easy maintenance and overhaul.
The door closing elastic can be adjusted, and the integrally formed silicone rubber door sealing ring ensures the height inside the box.
Storage, heating, testing, and drying are all carried out in an oxygen-free or inert atmosphere, so they do not oxidize.
Electric blast drying oven control execution system:
Temperature controller adopts touch button, digital display LED display, PID intelligent control instrument
Electric blast drying oven protection system:
The whole machine protection system consists of over temperature protection and alarm device.
The integrity of the components and test pieces is guaranteed.
Product introduction
Electric blast drying oven is widely used in the electromechanical, chemical, plastic, light industry and other scientific research units to bake, dry, solidify, heat treat and other convenient heating of various products and samples. Not applicable to volatile materials to avoid explosion.
Electric blast drying oven performance characteristics
1) 101 series electric blast drying oven is suitable for measuring moisture in coal , drying articles, drying heat treatment and other heating.
2) This series of high-precision digital temperature controller is used for temperature control, temperature control, easy operation and reliable performance. The digital display directly shows the working temperature and is easy to read.
The temperature of the studio can be arbitrarily constant between room temperature and 300 °C; it can automatically cut off the power supply and protect the equipment and samples.
This high-preparation is an indispensable analytical device in the laboratory.
Technical Parameters
Lu model 101-1A 101-2A 101-3A
Power (KW) 2 2.4 4
Voltage 220V 220V 380/220V
Temperature adjustment range from room temperature to 300 °C
Constant temperature fluctuation ± 1 ° C
Studio size (mm) 350*450*450 450*550*550 500*600*750
Dimensions (mm) 760*530*670 860*630*700 1250*840*640
Weight 50 60 75
Use and precautions
1. After connecting the power supply, you can turn on the 2 sets of heating switches, first turn on the blower switch to make the blower work, and then set the temperature of the control instrument to the temperature you need.
2. When the temperature rises to the desired temperature, the indicator light is off. At the beginning of the constant temperature, the temperature may continue to rise. This is the effect of residual heat, which will tend to stabilize in about half an hour. In the constant temperature process, the temperature control is automatically controlled by the thermostat in the box without manual management.
3. When the temperature is constant, a set of heating switches can be turned off, leaving only one set of electric heaters to work, so as to avoid excessive power and affect the sensitivity of the constant temperature.
4. The box should be placed indoors.
5, the supply lines should be installed in the knife switch an iron shell, for this special case, and the case ground.
6. Please check the electrical performance of the box before powering on, and pay attention to whether there is open circuit or leakage.
7. When everything is ready, put it into the sample, close the box door and unscrew the top exhaust valve. At this point, you can connect the power supply and start working.
8. Do not arbitrarily remove the side door, disturb or change the line. Only when the box fails, the side door can be removed and checked one by one according to the line. If there is a major fault, you can contact the factory.
9. This box is a non-explosion-proof drying box, so it has flammable volatiles. Do not put it in the dry box to avoid explosion.
10. Each drying oven is provided with two specimen shelves. The average load per shelf is 15 kg. Do not over-tighten and overload the test specimens to avoid affecting hot air convection. At the same time, the test piece should not be placed on the heat sink on the bottom of the studio to prevent overheating and damage the sample.
If the test pen or power adapter is damaged and needs to be replaced, you must replace the test pen and power adapter of the same model and the same electrical specifications. Do not use the instrument when the battery indicator indicates that the power is exhausted. If you do not use the instrument for a long time, remove the battery and store it.
Do not store or use the instrument in high temperature, high humidity, flammable, explosive or strong electromagnetic fields.
Use a damp cloth or detergent to clean the case. Do not use abrasives or solvents.
When the instrument is wet, please dry it and store it.
Electric blast drying oven operating procedures
(1) Dry box is used for baking, drying or other heating of the sample. The maximum operating temperature is 300 °C. The drying oven works under the conditions that the ambient temperature is not more than 40 ° C and the relative humidity of the air is not more than 85%.
(2) Use a special plug and socket and ground it with a wire that is twice as thick as the power cord. Check the electrical insulation performance before use, and pay attention to whether there is open circuit, short circuit and leakage.
(3) Put the thermometer on the box, turn on the power supply and adjust the temperature control knob to the set temperature. When the AC contact switch is just disconnected, check whether the reading of the thermometer matches the set value. If there is any difference, fine-tune it until the constant temperature is met. Set the temperature.
(4) Open the top or bottom vent to remove moisture from the box.
(5) The maximum average load of the test steel plate is 15Kg, and the placement should not be too heavy or too dense. Be sure to leave a gap. The test piece should not be placed on the bottom plate of the studio.
(6) It is strictly forbidden to put flammable and volatile substances in the drying box to prevent explosion.
(7) Connect the power supply, the green indicator light is on, the blower switch is turned on, the blower motor is running, the heating power is turned on, and the drying box is put into working state.
(8) When working, the door should not be opened frequently to avoid affecting the constant temperature field.
(9) There is no over-temperature protection device for this equipment. Therefore, there should be a special person to monitor the temperature inside the box when working. If the temperature is out of control, it should be checked off in time to avoid accidents.
solder ring Bronze Fittings produced as kembla and connex,the lead of the solder is much lower than them,solder ring bronze fittings,pre-solder bronze fittings,yorkshire solder ring fittings,connex solder ring fittings,solder ring gunmetal fittings,solder ring bronze fittings,pre-solder bronze fittings,yorkshire solder ring fittings,connex solder ring fittings,solder ring gunmetal fittings